Double shear load cell DT1


Download PDF

Double shear load cells for balances and scales, tension measure etc.

Double shear load cell

Upon request with ATEX or IECEx certification


Zoom Click to enlarge image

Double shear load cell DT1 Double shear load cell DT1 Schema
Analog Output ATEX IECEX

Technical Specifications

MaterialAluminium
Nominal load1000, 2500, 5000 kg
Limit load150 % F.S.
Breaking load300 % F.S.
Supply voltageMaximum 15 V dc
Output2 mV/V
Output tollerance0,1 % F.S.
Zero tollerance1 % F.S.
Linearity0,1 % F.S.
Hysteresis0,1 % F.S.
Reapeatability0,1 % F.S.
Input resistance2000-2050 ohm
Output resistance2000-2010 ohm
Insulation resistance>= 2000 Mohm
Creep (30 Minutes)0,1 % F.S.
Thermal compensation-10 / +40°C (14 / +104°F)
Operating temperature-20 / +70°C (-4 / +158°F)
Temperature deviation zero0,003 % F.S. / °C
Temperature deviation full scale0,002 % F.S. / °C
Protection classIP 67
Standard cable length1 m – 4 x 0,25 (shield not connected to the load cell body)
For the complete and latest technical specifications, please refer to the technical data sheet which can be downloaded here:
Download PDF

Connections

TypeSHIELDED CABLE 4C R5 PUR 4X0,25
Length1 m (body lead cell not connected shield)
Red+ IN
Black- IN
Green+ OUT
White- OUT
Connections  Double shear load cell DT1

MOUNTING OF THE LOAD CELL

General precautions to be followed in the installation and assembly of load cells:

  • Follow the direction of application of force on the load cell
  • Do not exceed the limits of the temperature as indicated in the data sheet
  • Do not remove labels from the load cell
  • Adequately protect the cables with sleeves or channels
  • Carry out on each load cell a ground connection with suitable cable, shunt on mounting accessories
  • Do not carry out any welding on the mechanical structure after mounting the load cell
  • Avoid or reduce to a minimum level any mechanical constraints between the rigid structure and the weighing structure to avoid error in weighing and return to zero
  • In case of vibration or dynamic loads product the load cells with puffers / shock absorbers
  • Protect load cell from any dust accumulation with any appropriate means